Skipt to content

Mineralogy & Geochemistry Laboratories Synthesis methods Berghof autoclave system & Teledyne ISCO syringe pump

Synthesis methods: Berghof autoclave system & Teledyne ISCO syringe pump

The new autoclave system from Berghof is being used to investigate the geochemical interaction of potential reservoir rocks with saline deep waters in conjunction with contaminated, supercritical CO2 as part of the COORAL joint project. In addition to the two Berghof autoclaves, this system includes a piston syringe pump from Teledyne ISCO and a measuring system for pH, eH and conductivity from Dr Michael Ecker Scientific Consulting.

  • Number of autoclaves 2 (type BR 2000)
  • Volume: 2 litres
  • Pressure: 200 bar max.
  • Temperature: 230 °C max (with PTFE insert)
  • Pressure vessel & lid stainless steel 1.4571 with PTFE insert/lining
  • Fittings 3 valves (stainless steel 1.4571), immersion tube made of PTFE (nominal diameter 4mm), temperature sensor in PFA-coated immersion tube, pressure gauge 250bar, monel bursting disc (200bar with PTFE foil)
  • Seals FPM (Viton) & PTFE
  • Dimensions inside 82 mm (without PTFE liner); outside 117 mm depth 345 mm (2000 ml)

  • Type: 500D
  • Volume 507ml l
  • Pressure 259bar
  • Delivery rate 1.0µl/min – 204ml/min
  • Resolution 31.7nl
  • Filling time 2.5min
  • Pressure control 10kPa (0.1bar)
  • Temperature 100°C max.
  • Additional pump equipment MLU
  • Electric inlet and outlet valve
  • Piston backflushing
  • Cooling – heating jacket
  • Possible operating modes of the pump
  • Pumping of liquids, gases and fluids
This translation in English is 3 days and 1 hours older than the original version (Deutsch) and may be slightly outdated. → View original in Deutsch if you are looking for any new developments.